EBV-Q Real-Time PCR Kit

Qualitative Detection of Epstein-Barr Virus by Real-Time PCR

The EBV-Q Real-Time PCR Kit enables qualitative detection of Epstein-Barr Virus (EBV 1 & 2) in human clinical samples using real-time PCR. It supports rapid and reliable identification of EBV DNA to aid in clinical diagnosis and monitoring.

At a Glance

Product Overview


The EBV-Q Real-Time PCR Kit is a probe-based molecular assay designed for the detection of Epstein-Barr Virus DNA in human samples. The assay utilizes specific primers and fluorescent probes for EBV detection, along with an internal control to ensure assay integrity and monitor amplification performance.

EBV-Q

How the Assay Works


  • Sample Collection & Nucleic Acid Extraction: DNA is extracted from plasma, serum, or CSF samples using validated extraction methods.
  • Real-Time PCR Amplification: Extracted DNA is amplified using EBV-specific primers in a single-tube reaction.
  • Fluorescent Detection: Amplification signals are detected in amplification signals are detected in real-time across fluorescence channels.
  • Result Interpretation: Ct value ≤ 35 is considered positive.

Key Features & Benefits


  • Single-tube assay design for simplified workflow
  • Specific detection of EBV DNA using probe-based chemistry
  • Integrated internal control for assay validation
  • Rapid and reliable detection
  • High sensitivity and specificity

Why EBV Detection Matters


Epstein-Barr Virus is one of the most common human viruses and is associated with infectious mononucleosis and other clinical conditions. Molecular detection using Real-Time PCR enables:

  • Rapid and sensitive detection of EBV DNA
  • Early identification of infection and reactivation
  • Improved diagnostic accuracy compared to conventional methods
  • Support for monitoring in immunocompromised patients

Sample Type & Reporting Output


Sample Types
  • Plasma, Serum, CSF
Reporting Output
  • Qualitative detection (Positive/Negative for EBV)

Workflow Advantage


Single-tube assay reduces handling steps and simplifies workflow
Pre-optimized master mix ensures consistent amplification performance
Internal control integrated within the assay enables reliable result validation
Closed-tube detection minimizes contamination risk
Compatible with standard real-time PCR instruments for routine laboratory use

Analytical Performance


The assay demonstrates reliable multiplex detection performance:

Analytical Sensitivity:
100%
Trueness (Bias):
99%
Limit of Detection (LoD):
1100 copies/ml
Measuring Range:
1 – 10⁹ copies/reaction
Clinical Sensitivity:
96%
Clinical Specificity:
99%
No cross-reactivity observed with HSV-1, HSV-2, Cytomegalovirus, and HIV

Storage and Stability


  • Shelf life: 12 months from date of manufacture
  • Storage: -20°C (shipped at -30°C to -10°C)
  • Avoid repeated freeze-thaw cycles (>6 times)
  • Protect from direct sunlight

Applications & Use Scenarios


The kit is intended for use in clinical laboratories for:

  • Detection of Epstein-Barr Virus infection
  • Monitoring EBV-associated conditions
  • Routine molecular diagnostic workflows
  • Support for clinical decision-making

Ordering Information


Commercial Name Old Cat No. New Cat No. Pack Size
EBV-Q Real-Time PCR Kit G2M705621 50 Tests

Product Enquiry Form


Since its inception in 2016, Genes2me has been constantly striving towards setting a benchmark in the diagnostics space by introducing premium quality (Made in India) diagnostic kits which are CE-IVD, ISO-13485:2016, and ISO 9001:2015 certified, assuring our clients of unparalleled quality and compliance with international standards.


© 2025 Genes2me. All rights reserved.